4 Apparatus and auxiliary aids
NOTE An instrument or item of apparatus is listed only if it is special, or made to a particular specification, usual laboratory glassware and equipment being assumed to be available.
4.1 Gas chromatograph
Equipped with a mass selective detector.
4.2 Gas chromatograph analytical capillary column
Appropriate operating conditions shall be established for the specific equipment used for the determination.
An example of suitable chromatographic conditions is given below:
Column 5 % phenylmethyl siloxane 30 m x 0,25 mm ID 0,25 μm film thickness
Carrier gas: Helium, 1 ml/min
Temperature program: isothermal at 70 "C for 2 min, from 70 °C to 250 °C at 10 °C/min
isothermal at 250 °C for 15 min
Injector: Splitless. 240 °C
Transferline: 280 °C
Detection: MSD (Mass Selective Detector)
lons, disopropy|naphthalene (DIPN) m/z 155, 197, 212
diethylnaphthalene (DEN) m/z 169
The following retention times have been observed under these conditions:
-- disopropylnaphthalene(DIPN) 6 major isomers 15,5 min to 16,1 min;
-- diethylnaphthalene (DEN) 14,8 min.
A separation of diisopropyinaphthalene (DIPN) into 7 isomers was observed with other columns, which did not influence the result.
4.3 Ultrasonic bath
4.4 Microsyringes, 100 μl, 250 μl
4.5 Graduated pipettes, 1 ml, 2 ml, 10 ml
4.6 40 ml screw top vials, with PTFE (polytetrafluorethylene) lined septa