Testing facility for determination of total in-can formaldehyde according to VdL-Guideline 03
Posted Date:2025/11/25
5.2.Acid steam distillation
(Determination of total in-can formaldehyde)
–Steam distillation apparatus, e.g. Vapodest 20s (Gerhard brand)
–Volumetric flasks 250 ml
–Precision balance, resolution of at least 0.001 g
–Spectrophotometer, suitable for a wavelength of 412 nm
–Graduated cylinder, 25 ml, with ground joint
–Suitable pipettes
–Suitable cuvettes
- Sulphuric acid 25% (v/v),
e.g. made up from 839 ml deionised water and 160 ml conc. sulphuric acid
–Acetyl acetone reagent for a volume of 500 ml aqueous solution:
–75 g ammonium acetate p.a.
–1 ml acetylacetone (2,4-pentandione) p.a.
–1.5 ml acetic acid 100% p.a.
–Fill up to 500 ml with deionised water
–Deionised water
Note:
The acetyl acetone reagent is not stable. If kept at room temperature, it may be used for testing for no longer than 14 days.